Failures of Heating Cooling Circulators
Il heating cooling circulator can be used in many chemical pharmaceuticals and laboratories. Then, how to solve the failures of the heating cooling circulator?
Failure of circulating pump
The failure of the circulating pump is generally the heat overload of the circulating pump. There are 4 main reasons for this failure:
1. The external circulation pipeline is too long. It is suggested to shorten the external circulation pipeline, reduce the bending of the external circulation pipeline, and then click the Reset button of the electric cabinet surface heat relay;
2. The circulating pipe is blocked. It is suggested to check the degree of cleanliness of the circulating pipe, install a filter, and then click the Reset button of the electric cabinet heat relay;
3. The circulation pump voltage is too low. It is suggested to measure whether the circulation pump voltage is in line with the prescribed range, improve the voltage environment, and then click the Reset button of the electric cabinet heat relay;
4. The regulating current of the heat relay is too small. It is suggested to measure the pump current, compare the heat relay range, turn up the heat relay current properly with the cross screwdriver, and then click the Reset button of the electric cabinet heat relay.
Failure of sensor
The great possibility of sensor failure is the wiring error. At this time, check the wiring in time, and we should pay attention to two kinds of situations, one is the sensor is short-circuited or the temperature is < -160 ℃, the other is the sensor is open circuited or the temperature is > 349℃.
If the sensor is short-circuited or the temperature is <-160 ℃, it may be the fault of the sensor itself. At this time, change the sensor (there is a standby sensor in the equipment line slot), if the sensor connection is wrong, it is suggested to check the connection condition (the control circuit diagram) to correct.
If the sensor is open circuited or the temperature is > 349 ℃, it should be the fault of the sensor itself. At this time, replace the sensor in time (there is a standby sensor in the equipment line slot). It is necessary to pay attention to whether the temperature is too high and beyond the range. If the equipment is out of electricity, check the power supply of the equipment.
The above two failures of the heating cooling circulator are relatively simple. I believe that you can quickly solve these failures with these tips.
Raccomandazioni correlate
-
Beginning with Refrigeration, Heating and Temperature Control System Product Diversification Develo
997Wuxi Guanya Refrigeration technology co., ltd was founded in 2006. It set up a refrigeration and heating equipment studio. In 2007, it launched the first generation refrigeration and heating cycle, registered the trademark "LNEYA". In 2008, it lau...
Visualizza i dettagli -
Application area of high temperature refrigeration and cooling control system
850The high-temperature refrigeration and cooling control system is used for supporting glass reactors, rotary evaporators, fermentation tanks, and calorimeters; widely used in petroleum, chemical, metallurgical, biochemical, physics, testing and che...
Visualizza i dettagli -
Refrigeration heating circulator list resonance fault description
1183Refrigeration heating circulator leaderboard works: When the temperature of the reactor, reactor, rotary evaporator, extraction equipment, etc. of the refrigeration heating circulator needs to be rapidly increased or lowered, the equipment can qui...
Visualizza i dettagli -
Selection of Temperature Control System for Laboratory Resin Reaction Kettle
10141. Choose from the quality of the laboratory resin reactor temperature control system: can affect the quality of the laboratory resin reactor temperature control system, may be the use of accessories. There are some small factories on the market t...
Visualizza i dettagli