Common Faults of Salt Spray Environmental Chambers
The enterprise users are often at a loss when dealing with common faults in the salt spray environmental chamber. For this reason, Wuxi LNEYA, a manufacturer of the salt spray environmental chamber, specially reorganized the related common fault treatment technologies of the salt spray environmental chamber.
Salt spray environmental chamber for sale
The salt spray environmental chamber cannot be heated
1. Heat the solid state relay, if there is no output, replace the solid state relay.
2. If the external dial has no output, replace the external dial.
3. If there is heating tube damage, check the heating tube resistance, replace the resistance if there is something abnormal.
4. Lack of water can cause heating shielding, check whether the water level is normal or not and whether the floating ball malfunctions or not.
5. If there is a malfunction of the over-temperature maintenance system, adjust the maintenance temperature value. If the fault still cannot be solved, replace the over-temperature maintenance system.
The heating temperature is abnormal
1. If the set temperature of the temperature controller of the environmental chambers for sale is too low, set the temperature controller at the required temperature.
2. If the set temperature of the safety maintenance switch of the environmental chamber is too low, set the safety maintenance switch at the required temperature.
3. If the set temperature of the temperature controller of the full bucket is too low, set the temperature controller at the required temperature.
4. If the set temperature of the safety maintenance switch of the full bucket is too low, set the safety maintenance switch at the required temperature.
The external dial doesn’t display
1. Check whether or not there is a power supply in the salt spray environmental chamber. If not, ask a professional electrician for a check.
2. Check whether or not the power switch is damaged. If damaged, replace the power switch.
3. Check whether or not the power circuit breaker is damaged. If damaged, replace the circuit breaker.
4. Check whether or not the external dial terminal has a power supply. If there is a power supply, it indicates that the external dial is damage, replace the external dial.
5. Check whether or not the power supply wiring terminal of the salt spray environmental chamber is blackened. If it is blackened, it indicates the possibility of burning out, replace the external dial.
If the failures the enterprise users are encountered are more serious than above, it is suggested to contact the salt mist environmental chamber manufacturer to solve the failure.
Raccomandazioni correlate
-
Several Common Troubleshooting Methods for Ultra-low Temperature Freezer in Laboratories
16321. After power supply is switched on, the air switch of the main power supply is automatically disconnected.Elimination method: Check whether the compressor of refrigeration system is damaged or not, and replace the parts if damaged. After replaci...
Visualizza dettagli -
Considerazioni sulla scelta del sistema di controllo della temperatura del sistema di distillazione
1990Lneya distillation system temperature control system is a self-developed refrigeration heating temperature control system for the distillation industry supporting equipment, which is highly sought after in the distillation industry. So what should...
Visualizza dettagli -
Qual è la gamma di controllo della temperatura dei refrigeratori industriali?
2447Il refrigeratore industriale è un'apparecchiatura di refrigerazione comune nell'industria della refrigerazione industriale, la cui caratteristica è costituita da vari tipi, modelli completi e un'ampia gamma di applicazioni.
Visualizza dettagli -
Battery test chiller operation evaporating temperature description
1866In the current new energy vehicle battery test equipment, the use of more batteries to test the chiller evaporation temperature has a certain impact on the equipment, then what is the specific impact? When the battery test chiller heat load increa...
Visualizza dettagli
Refrigeratori industriali LNEYA Produttore Fornitore